Back to Comparisons
Product Comparison

Resound Vivia 9 60S microRIE vs Signia Active Mini IX

Side-by-side HearAdvisor lab results with SoundGrades, expert pick badges, and the same visualizations we use in product reviews.

Uses Official HearAdvisor Lab Data
Resound Vivia 9 60S microRIE Hero
ResoundSoundGrade B

Resound Vivia 9 60S microRIE

HearAdvisor lab-tested performance overview.

Expert Choice
Expert
Choice
Product Image
SigniaSoundGrade A

Signia Active Mini IX

HearAdvisor lab-tested performance overview.

We independently review each item we recommend and appreciate your trust and support. When you buy through our links, we may earn a commission. Learn More.

Performance metrics

Scale 1-5 (Tuned)

Lab metric by fit

Scale 1-5 (Tuned)

Speech in quiet
Resound Vivia 9 60S microRIE4.0
Signia Active Mini IX4.2
Speech in noise
Resound Vivia 9 60S microRIE1.3
Signia Active Mini IX2.9
Own voice
Resound Vivia 9 60S microRIE3.4
Signia Active Mini IX3.2
Feedback handling
Resound Vivia 9 60S microRIE4.2
Signia Active Mini IX2.5
Music streaming
Resound Vivia 9 60S microRIE2.6
Signia Active Mini IX3.6
Overall fit score
Resound Vivia 9 60S microRIE4.1
Signia Active Mini IX4.5

Speech level

Adjust input dB (Tuned)

Amplification (REIG)

Insertion gain at 65 dB (Tuned)

Fitting configuration for tuned fit

Resound Vivia 9 60S microRIESignia Active Mini IX
Ear CouplingTulip Dome
Ear CouplingSleeve Vented
ReceiverMedium-power
ReceiverIntegrated in device
Fitting FormulaAudiogram+ (Manufacturer proprietary targets)
Fitting FormulaIXFit (Manufacturer proprietary targets)
Listening ProgramAll-Around (Automatic)
Listening ProgramUniversal (Automatic)
Volume SettingAdjusted to match NAL-NL2 targets at 55, 65, and 75 dB SPL
Volume SettingAdjusted to match NAL-NL2 targets at 55, 65, and 75 dB SPL
Software VersionSoftware 2.1.305.0 Device Firmware v10.21.2.1
Software Versionv25.7.1142.3
Tuning AdjustmentsProgrammed to N3 audiogram, Feedback calibrated, Experience Level set at "Experienced Non-Linear", Gain % set at "100%", Other settings at manufacturer defaults
Tuning AdjustmentsProgrammed to N3 audiogram, Feedback calibrated, Experience Level set at "Experienced", Other settings at manufacturer defaults
Notesn/a
Notes